标准编号:EN50305, BS6853附录B
应用领域:
检测目的:Smoke Toxicity
行业标准: EN European standard BS British Standards Institute
The Toxicity Index Tester is used for flue gas toxicity tests such as EN 50305 and BS 6853 Appendix B. During the test, a certain flow of flue gas is sent into a standard adsorption liquid for absorption, or collected in a gas collection bag for the next step of analysis and use.
EN 50305: British Cable Standard
BS 6853 Appendix B: Toxicity Test Criteria
ISO 19700: Toxicity testing of fire air streams - Steady-state tube furnace test method
Equipment No.: PX01011
Equipment Dimension:3600(W)×900(D)×1800(H)mm
Gas Analysis Cabinet Dimensions:650(W)×680(D)×1800(H)mm
Power Supply: AC 220V,25A
Weight: APPR.550kg
Gas Source: Compressed Air
Sealed box, the flue gas generating device is installed in the box, the interface is clean, and the operation of the equipment can be well protected; the large observation window is convenient for the operation of the equipment.
The structure, appearance and interface design of imported industrial design concepts, integrated panel structure, convenient operation, friendly interface, ergonomic and operating habits, beautiful appearance
Tube furnace heating, the length is 600mm, automatic adjustment makes the furnace body evenly heated to the target temperature; the tube furnace is protected by a mesh shell, which can effectively dissipate heat; the high temperature K-type thermocouple measures the temperature of the sample area, and the measurement range is 0-1100℃ , the measurement accuracy is ±0.1℃.
The quartz tube made of heat-resistant quartz above 2mm is used for gas transmission and sample heating, and the pipe joint is frosted to increase the sealing performance
The ceramic combustion boat is controlled to move by a strong magnet and a transmission bar, which can stably transfer the measured material to the furnace core measurement area
The screw drive mechanism can move the traction tube to the quartz seal, with accurate positioning and stable operation to avoid damage to the quartz tube caused by human factors
Equipped with two filter columns, packed with activated carbon and color-changing silica gel, to clean and dry the carrier gas
The non-continuous analysis system for toxic gas components is equipped with imported gas collection bags and imported colorimetric tubes; SO₂: measuring range: 0-1000ppm, resolution 1ppm, accuracy 30ppm; NOX: measuring range: 0-3000ppm, resolution 1ppm, accuracy 30ppm
Gas sampling bag with PTFE coating, capacity 80L, equipped with imported manual sampler
Optional online infrared gas analyzer and its gas pretreatment system:
CO: Measuring range: 0~10%; Resolution: 0.01%, Accuracy: Absolute ±0.1% or relative 3%.
CO₂: measuring range: 0~20%; resolution: 0.01%, accuracy: absolute ±0.2% or relative 3%
Toxic gas discontinuous analysis system: (UV spectrophotometer) (optional)
HCN: Spectral analysis equipment: Spectral analyzer with wavelength of 500nm, permeable membrane thickness 10mm
Reagent: Gas bubbling absorption solution is 0.1N NaOH
Titration reagent: aqueous solution of picric acid with a concentration of 3g/dm3 and aqueous solution of sodium carbonate with a concentration of 50g/dm3
Brand computer + professional operating software to automatically control the test process
Temperature control and stabilization are automatically controlled by software, recording and curve formation
Test data is automatically displayed, stored and reported