在线客服 电话 微信 置顶
电话:400-086-0699

微信客服

微信客服

首页>PRODUCT > Flammability>Needle flame combustion tester

PX03004 Needle flame combustion tester

产品型号:PX03004

标准编号:GB/T5169.5-2008, IEC60695-11-5

应用领域: 电线电缆 其他领域

检测目的:Flame SpreadFlammability

行业标准: GB中国标准 IEC国际电工委员会

Needle flame combustion test tester, designed for under specified conditions the experimental flame will not ignite parts, or the experimental flame ignites combustible parts, but the combustion duration or combustion length of parts is limited, and the flame or burning or hot particles falling from the test sample will not be the test equipment for combustion spread. High degree of automation, accurate test data.

产品标准产品规格产品特点使用单位

产品标准

GB/T 5169.5 Experimental flame needle flame test method apparatus, validation test method and guidelines


Iec60695-11-5 Needle flame test method. Protocols and guidelines for installation and verification tests


产品规格

Equipment No.:                         PX03004

Dimension:                             1350(W)×700(D)×1000(H)mm  

Bottom Chamber Dimension:1350(W)×700(D)×650(H)mm

Power Supply:                           AC 220V, 50/60Hz, 2A

Weight:                                      APPR.70kg

Gas Source:                               Industrial methane with purity above 98%


产品特点

  • Steel structure box, test box volume more than 0.75 cubic meters, to provide adequate clean air for the test.


  • Large observation window, the inner wall of the box is coated with dark color, and the illumination is within 20Lux, which is convenient to observe the combustion condition of the sample.


  • Smoke exhaust device is installed on the top, which can automatically open to discharge test exhaust gas after the end of the test.


  • Introduction of industrial design concept of structure, appearance and interface design, integrated panel structure, easy to operate, friendly interface, ergonomic and operating habits, beautiful appearance


  • The burner made of brass has good thermal stability and low deformation at high temperature. The diameter of the fire nozzle is 0.5±0.1mm, the outer diameter is 0.9mm, and the length is 35mm. It meets the standard requirements and has a protective device to prevent accidental damage.


  • The Angle of the burner can be easily adjusted to 0 or 45 degrees to meet the position requirements of the sample subjected to fire and temperature correction


  • High pressure pulse automatic ignition, ignition stability and reliability, high security


  • Equipped with stainless steel fixture, durable; The sample fixture can achieve smooth movement and precise positioning in horizontal and vertical directions by sliding rail structure.


  • Equipped with imported precise rotor flowmeter to control gas flow, equipped with imported differential pressure measurement burner back pressure.


  • Equipped with brand pressure gauge and pressure regulating valve to adjust gas pressure, equipped with solenoid valve to control gas on and off, to ensure the safety of equipment.


  • Programmable logic controller (PLC)+ touch screen control, can realize automatic control/detection/calculation/data storage


  • Test time, fire time can be set by oneself, software automatic timing


  • Linear combustion rate (V) calculated by PLC, data displayed and saved by touch screen


  • It has the function of reminding the arrival of flame impact time and automatically cutting off the air source without experiment for 10 minutes to ensure the safety of use.


  • Optional temperature calibration package, imported thermocouple measurement accuracy ±0.1℃, can automatically generate calibration curve and determine.

使用单位